Emerson Micro Motion 2700 Configuration And Use Manual page 37

With profibus-pa
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Calibration
Table 3-2
PROFIBUS parameters for Smart Meter Verification
Step number
1
2
3
4
5
6
7
8
Configuration and Use Manual
Step description
Set output state
Start/abort procedure
Check current algorithm state
Read percent complete
Check algorithm abort state
Check inlet stiffness
Check outlet stiffness
Read abort code
Parameters
Block: Transducer block 1
Index: 182
Value:
• 0: Last measured value (default)
• 1: Fault
Block: Transducer block 1
Index: 72 (Start/Stop Meter Verification)
• 0x00: No effect
• 0x01: Start On-Line Meter Verification
Block: Transducer block 1
Index: 75
Value:
• Bits 4–6: State
Block: Transducer block 1
Index: 189 (Progress)
Block: Transducer block 1
Index: 75
Value:
• Bits 0–3: Abort code
Block: Transducer block 1
Index: 77
• 0: Within uncertainty limit
• 1: Outside uncertainty limit
Block: Transducer block 1
Index: 78
• 0: Within uncertainty limit
• 1: Outside uncertainty limit
Block: Transducer block 1
Index: 185
Codes: See Table 3-3
29