Operation
With Prolink II
To control concentration measurement totalizers and inventories, choose
Control
. To control all other totalizer and inventory functions, choose
To reset inventories using ProLink II, you must first enable this capability. To enable inventory reset
using ProLink II:
1. Choose
View > Preferences
2. Select the
3. Click
Apply
With PROFIBUS EDD
To start and stop totals and inventories, to reset all totals and inventories simultaneously, or to reset
individual mass or volume totals and inventories, select
To reset petroleum measurement totals and inventories alone, select
To reset concentration measurement totals and inventories alone, select
.
Totalizer
With PROFIBUS bus parameters
If the totalizer blocks are configured to report the status of one of the internal totalizers (i.e., not
Standard mode) (see Section 2.6), you can reset that total or inventory by setting index 29 of the
appropriate totalizer function block to 1.
You can also control the internal totalizers directly with the transducer block parameters shown in
Table 5-3. Except where specified, enabling each function means setting its value to 0x0001.
Table 5-3
Stopping, starting, and resetting totalizers and inventories
To accomplish this:
Stop all totalizers and inventories
Start all totalizers and inventories
Reset all totals
Reset all inventories
Reset mass total
Reset mass inventory
Reset liquid volume total
Reset liquid volume inventory
Reset gas standard volume total
Reset gas standard volume inventory
Reset petroleum measurement volume totalizer
Reset petroleum measurement volume inventory
Reset ED standard volume total
Reset ED standard volume inventory
Reset ED net mass total
Reset ED net mass inventory
Reset ED net volume total
Reset ED net volume inventory
Configuration and Use Manual
.
Enable Inventory Totals Reset
.
ProLink > CM Totalizer
ProLink > Totalizer Control
checkbox.
Device > Device > Totalizer
Device > Device > API Totalizer
Device > Device > CM
Use this transducer block parameter:
Slot
Index
11
49 (value = 0x0000)
11
49 (value = 0x0001)
11
50
11
51
11
52
11
60
11
53
11
61
11
70
11
71
12
36
12
37
12
60
12
63
12
61
12
64
12
62
12
65
.
.
.
91