Calibration
Figure 3-12
D1 and D2 density calibration – EDD
Calibration >
Density Cal
D1 calibration
Fill sensor with D1
fluid
D1 = density of D1
fluid
Start Lo Density Cal
Execute
Low Density Cal in
progress
Figure 3-13
D1 and D2 density calibration – Bus parameters
D1 Calibration
Fill sensor with D1
fluid
Enter density of D1
fluid
Initiate D1 calibration
Check status
Check for failure
alarms
Check K1 value
Configuration and Use Manual
D2 calibration
Fill sensor with D2
D2 = density of D2
Start Hi Density Cal
High Density Cal in
Block: Transducer Block 1 (Slot 11)
Index 97 (D1)
Block: Transducer Block 1 (Slot 11)
Index 87 (low density cal)
Block: Transducer Block 1 (Slot 11)
Index 143, Bit 0x4000
Block: Transducer Block 1 (Slot 11)
Index 141, Bits 0x0100, 0x0200,
and 0x0400
Block: Transducer Block 1 (Slot 11)
Index 92 (K1)
fluid
fluid
Execute
progress
Done
D2 Calibration
Fill sensor with D2
fluid
Enter density of D2
fluid
Initiate D2 calibration
Check status
Check for failure
alarms
Check K2 value
Done
Block: Transducer Block 1 (Slot 11)
Index 98 (D2)
Block: Transducer Block 1 (Slot 11)
Index 88 (high density cal)
Block: Transducer Block 1 (Slot 11)
Index 143, Bit 0x2000
Block: Transducer Block 1 (Slot 11)
Index 141, Bits 0x0100, 0x0200,
and 0x0400
Block: Transducer Block 1 (Slot 11)
Index 93 (K2)
41