Calibration
Figure 3-18
Temperature calibration – ProLink II
Temperature Offset calibration
Fill sensor with low-temperature fluid
Wait until sensor achieves thermal
ProLink Menu >
Calibration >
Temp offset cal
Enter temperature of low-temperature
Calibration in Progress light turns red
Calibration in Progress light turns
Configuration and Use Manual
equilibrium
fluid
Do Cal
green
Close
Temperature Slope calibration
Fill sensor with high-temperature fluid
Wait until sensor achieves thermal
equilibrium
ProLink Menu >
Calibration >
Temp slope cal
Enter temperature of high-temperature
fluid
Do Cal
Calibration in Progress light turns red
Calibration in Progress light turns
green
Close
Done