Model 2700 PROFIBUS Block Parameters
Table F-2
Physical block parameters (continued)
Index
Parameter Mnemonic
43
COND_STATUS_DIAG
44
Reserved
45
Reserved
46
Reserved
47
Reserved
48
Reserved
49
Physical Block Views
F.3.1

Physical block object

Table F-3 shows the physical block object.
Table F-3
Physical block object
Slot/Index
Element name
Slot 0/Index 16
Reserved
Block_Object
Parent_Class
Class
DD_Refrence
DD_Revision
Profile
Profile_Revision
Execution_Time
Number_Of_Parameters
Address_of_View_1
Number_of_Views
Configuration and Use Manual
Definition
Message
Type
Condensed Status Diagnostics
Simple
Data type
Unsigned 8
Unsigned 8
Unsigned 8
Unsigned 8
Unsigned 32
Unsigned 16
Octet String
Unsigned 16
Unsigned 8
Unsigned 16
Unsigned 16
Unsigned 8
Data Type/
Size
Store
Default
Structure
/Rate
Value
(HZ)
Unsigned-8
1
S
0
Size in bytes
Value
1
250 (default)
1
01
1
01
1
250 (default)
4
00 ,00, 00, 00 (Reserved)
2
00 ,00 (reserved)
2
64 02 (compact class B)
2
03 01 (3.01)
1
00 (for future use)
2
00 26 ( Max no of Physical Block
Parameters)
2
00 49 (slot, index)
1
01 (1 view)
Access
Enumerated List of
Modbus
Values / Range
Register /
Coil
R/W
0: Status and
NA
Diagnosis is
provided as
defined in
-PROFIBUS
Profile:
"PROFIBUS-PA
Profile for Process
Control Devices"
V3.01,
December 2004.
PNO-Order-No.
3.042.
1: Condensed
Status and
Diagnosis
information is
provided. NA