Configuration
4.17

Enabling LD Optimization

LD Optimization is a special compensation is that is specifically for hydrocarbon liquids. LD
Optimization should not be used with any other process fluids. LD Optimization is available only with
certain large sensor sizes. If your sensor can benefit from LD Optimization, the enable/disable option
will appear in ProLink II or on the display.
If you send the transmitter to a calibration facility to perform a water calibration, either during startup or
any time thereafter, LD Optimization must be disabled. When you have completed the calibration,
re-enable LD Optimization.
To enable LD Optimization, see Figures 4-21 and 4-22.
Figure 4-21
80
LD Optimization – ProLink II
Model 2700 Transmitter with PROFIBUS-PA