Calibration
Figure 3-15
D3 or D3-and-D4 density calibration – EDD
Calibration >
T-Series Density Cal
D3 calibration
Fill sensor with D3
fluid
D3 = density of D3
fluid
Start D3 Density Cal
Execute
D3 Cal in progress
Done
Figure 3-16
D3 or D3-and-D4 density calibration – Bus parameters
D3 Calibration
Fill sensor with D3
fluid
Enter density of D3
Block: Transducer Block 1 (Slot 11)
fluid
Index 100 (D3)
Block: Transducer Block 1 (Slot 11)
Initiate D3 calibration
Index 90 (D3 cal)
Block: Transducer Block 1 (Slot 11)
Check status
Index 143, Bit 0x0040
Block: Transducer Block 1 (Slot 11)
Check for failure
Index 141, Bits 0x0100, 0x0200,
alarms
Block: Transducer Block 1 (Slot 11)
Check K3 value
Index 95 (K3)
Done
Configuration and Use Manual
D4 calibration
Fill sensor with D4
fluid
D4 = density of D4
fluid
Start D4 Density Cal
Execute
D4 Cal in progress
Done
and 0x0400
D4 Calibration
Fill sensor with D4
fluid
Enter density of D4
Block: Transducer Block 1 (Slot 11)
fluid
Index 101 (D4)
Block: Transducer Block 1 (Slot 11)
Initiate D4 calibration
Index 91 (D4 cal)
Block: Transducer Block 1 (Slot 11)
Check status
Index 143, Bit 0x0080
Block: Transducer Block 1 (Slot 11)
Check for failure
Index 141, Bits 0x0100, 0x0200,
alarms
Block: Transducer Block 1 (Slot 11)
Check K4 value
Index 96 (K4)
Done
and 0x0400
43