Troubleshooting
8. Test the terminal pairs as follows:
Brown against all other terminals except Red
Red against all other terminals except Brown
Green against all other terminals except White
White against all other terminals except Green
Blue against all other terminals except Gray
Gray against all other terminals except Blue
Orange against all other terminals except Yellow and Violet
Yellow against all other terminals except Orange and Violet
Violet against all other terminals except Yellow and Orange
Note: D600 sensors and CMF400 sensors with booster amplifiers have different terminal pairs.
Contact Micro Motion Customer Service for assistance.
There should be infinite resistance for each pair. If there is any resistance at all, there is a short
between terminals.
9. See Table 6-13 for possible causes and solutions.
10. If the problem is not resolved, contact Micro Motion Customer Service.
Note: When reassembling the meter components, be sure to grease all O-rings.
Table 6-13
Sensor and cable short to case possible causes and remedies
Possible cause
Moisture inside the sensor junction box
Liquid or moisture inside the sensor case
Internally shorted feedthrough (sealed passage for
wiring from sensor to sensor junction box)
Faulty cable
Improper wire termination
6.14.2

4-wire remote or integral installation

If you have a 4-wire remote installation or an integral installation:
1. Power down the transmitter.
2. If the transmitter is in a hazardous environment, wait five minutes.
3. If you have a 4-wire remote installation, remove the core processor lid.
4. If you have an integral installation:
a. Loosen the four cap screws that fasten the transmitter to the base (Figure 6-1).
b. Rotate the transmitter counter-clockwise so that the cap screws are in the unlocked
position.
c. Gently lift the transmitter straight up, disengaging it from the base.
Note: You have the option of disconnecting the 4-wire cable or leaving it connected.
Configuration and Use Manual
Solution
Make sure that the junction box is dry and no corrosion is
present.
Contact Micro Motion Customer Service.
Contact Micro Motion Customer Service.
Replace cable.
Verify wire terminations inside sensor junction box. See the
Micro Motion 9-Wire Flowmeter Cable Preparation and
Installation Guide or the sensor documentation.
111