Changing The Measurement Mode Parameter - Emerson Micro Motion 2700 Configuration And Use Manual

With profibus-pa
Hide thumbs Also See for Micro Motion 2700:
Table of Contents

Advertisement

Configuration
4.14

Changing the measurement mode parameter

The measurement mode parameter defines how the flow is added to or subtracted from the totalizers.
Forward flow moves in the direction of the arrow on the sensor.
Reverse flow moves in the direction opposite from the arrow on the sensor.
Table 4-15 shows the possible values for the measurement mode parameter and the transmitter's
behavior when the flow is positive or negative. Only the unidirectional and bidirectional values are
recognized by the PROFIBUS specification, so other values will be unrecognized by a PROFIBUS
host or configuration tool. However, the transmitter will operate correctly in any of the modes listed in
Table 4-15.
Table 4-15
Totalizer behavior for each measurement mode value
Measurement mode value
Unidirectional (forward only)
Reverse only
Bidirectional
Absolute value
Negate/forward only
Negate/bidirectional
To change the measurement mode parameter, refer to the flowcharts in Figure 4-15.
Figure 4-15
Changing the measurement mode parameter
EDD
MMI Coriolis Flow >
Transducer Block >
Measurement >
Process Variable >
Mass Flow
Measurement Mode
Configuration and Use Manual
Bus index
Forward flow
0
Increase
1
No change
2
Increase
3
Increase
4
No change
5
Decrease
Bus parameters
Block: Transducer Block 1 (Slot 11)
Measurement Mode
Index 10 (measurement mode)
ProLink II
ProLink >
Configuration
Flow tab
Select a value from
the Flow Direction
list
Apply
Reverse flow
No change
Increase
Decrease
Increase
Increase
Increase
73