<Parameter explanation>
Parameter
Description
Save intervals
Sets the intervals(0 to 1000) to be internally logged.
If "1" is set , all measured data is stored, and "2" is set, one measured data is stored every two
measurement.
If "0" is set, only the applied measured data is stored when hold is set.
Save count
Sets the maximum data count (1 to 12,800) to be internally logged.
The internal logging process ends when the number of internal logging data reaches the maximum.
Internal logging end command <LE command>
End the internal logging of the data.
<Command format>
L
E
CR
<Response format>
Normal processing
O
K
CR
When a command was not successfully processed/When internal logging is not started
E
R
CR
• The internal logging process ends without sending LE command when the number of internal logging data
reaches the maximum.
Internal logging data acquisition command <LO command>
Acquires the internal logging data.
<Command format>
L
O
↑
Space
↑
Task number
• If the <task number> is omitted, internal logging data acquisition is executed for the task of which the result is
currently displayed.
• If the <first data number> is omitted, internal logging data acquisition is executed from first data number "0".
• If the <output data count> is omitted, all internal logging data acquisition is executed.
ZW User's Manual
↑
↑
↑
Space
Space
↑
First data number
CR
Output data count
No-protocol Connection
191