Omron ZW-C1*T User Manual page 173

Displacement sensor zw series confocal fiber type
Hide thumbs Also See for ZW-C1*T:
Table of Contents

Advertisement

Initial Settings for No-protocol Communications
You must set the communications method, destination IP address, and I/O port number of the destination
external device to perform no-protocol communications.
 Explorer pane
: [Device group] | Sensor name | [System] | [System data] (Double-click)
 Edit Pane
: [Ethernet communication settings] icon (
Item
Communication type
Port No. In
IP address
Port No. Out
Important
If set Communication type to TCP client, please start a ZW after make sure the external device to be TCP
server is ready to communicate. In addition, the communication is interrupted for reasons such as the ethernet
cable is disconnected, please restart the ZW If you can not reconnect.
ZW User's Manual
[No-Protocol data communication setting]
Description
Select the communications method.
Sets the ZW port number when the UDP or TCP server is
selected.
Set the IP address of the external device at the connection
destination when the UDP or TCP client is selected.
Set it in the form a.b.c.d.
Note
If you connect an external OMRON CS/CJ-
series PLC to Ethernet, the following default IP
address is assigned to the PLC.
• IP address: 192.168.250.node_address
Set the I/O port number of the external device at the
connection destination when the UDP or TCP client is
selected.
Set the value to between 0 and 65,535.
) |
Setting range
• OFF
• TCP server
• TCP client
• UDP
(Default: TCP server)
0 to 65,535
(Default: 9601)
• The following port number is reserved
and cannot be used: 9600
a: 1 to 223, b: 0 to 255,
c: 0 to 255, d: to 254
(Default: 192.168.250.100)
0 to 65,535
Default: 9,600
(Default: 9,601)
No-protocol Connection
171