Measured value 2 mode (When measured values for multiple tasks are output by BINARY_SEL
input)
Judgment output
(HIGH/PASS/LOW)
TASK A
measurement value
TASK B
measurement value
BINARY_SEL input
Binary output
ON
GATE output
OFF
BINARY OUT output
Item
T0
Measurement cycle
T4
GATE signal rise time
T5
GATE signal output time
T15
Binary output update cycle
T17
Response time of input
Explanation of operations (When the results for all tasks are output)
(1) Input the number of the task you want to output to the BINARY_SEL input.
(2) The results for the selected tasks are output (judgment output and binary output).
(3) After the judgment result and binary data output, the GATE output is turned ON.
(4) GATE signal output time after the GATE output is turned ON, the GATE output is turned OFF.
(5) When the results for the selected tasks are updated, the judgment output and binary output are also
updated.
(6) Input the number of the task you want to output to the BINARY_SEL input.
(7) The results for the selected tasks are output (judgment output and binary output).
(8) After the judgment result and binary data output, the GATE output is turned ON.
(9) GATE signal output time after the GATE output is turned ON, the GATE output is turned OFF.
(10) When the results for the selected tasks are updated, the judgment output and binary output are also
updated.
ZW User's Manual
(1)
(2)
(3)
(4) (5)
T15
T4
T17
T0
T5
TASK A
TASK A result
TASK A
Min.
0.5 ms
Setting value
(0.1 to 100.0 ms)
T0×Update cycle
(6)
(7)
(8)
(9)
T17 T4
T5
TASK B
TASK A result
TASK B result
TASK B
Max.
Depends on the set conditions
0.2 ms
Setting value
(0.1 to 100.0 ms)
T0×Update cycle
T0
Parallel I/O connection
(10)
TASK B result
167