Omron ZW-C1*T User Manual page 168

Displacement sensor zw series confocal fiber type
Hide thumbs Also See for ZW-C1*T:
Table of Contents

Advertisement

Task Measurement Result Output
The following two output methods can be selected from.
• Measured value 1 mode (When measured values for multiple tasks are output consecutively)
The preset tasks are output in order for each of a certain cycle.
• Measured value 2 mode (When measured values for multiple tasks are output by BINARY_SEL input)
The tasks to output with binary output are selected with external input and the measurement data for those tasks is
output.
Measured value 1 mode (When measured values for multiple tasks are output consecutively)
The binary output results can be acquired when the GATE output is ON.
Judgment output
(HIGH/PASS/LOW)
Binary output
GATE output
BINARY OUT output
Item
T0
Measurement cycle
T4
GATE signal rise time
T5
GATE signal output time
T15
Binary output update cycle
T16
Binary output update cycle 2
Explanation of operations (When the results for all tasks are output)
(1) The TASK1 results are output (judgment output and binary output).
(2) After the judgment result and binary data output, the GATE output is turned ON.
(3) The GATE signal output time after the GATE output is turned ON, the GATE output is turned OFF.
(4) After the binary output update cycle, the TASK2 results are output (judgment output and binary output).
(5) After the binary output update cycle, the TASK3 results are output (judgment output and binary output).
(6) After the binary output update cycle, the TASK4 results are output (judgment output and binary output).
(7) After binary output update cycle 2, the TASK1 results are output (judgment output and binary output).
Note
Which tasks to output results for can be selected.
Binary output object task selection input p.136
166
Parallel I/O connection
(1) (2)
T15
T0
T4
T5
TASK3 result
TASK4 result TASK1 result TASK2 result TASK3 result TASK4 result TASK1 result TASK2 result
ON
OFF
010b
011b
000b
(3)
(4)
(5)
(6)
T16
001b
010b
Min.
0.5 ms
Setting value
(0.1 to 100.0 ms)
T0×Update cycle
T0×Update cycle×number of tasks
output
(7)
011b
000b
001b
Max.
Depends on the set conditions
0.2 ms
Setting value
(0.1 to 100.0 ms)
T0×Update cycle
T0×Update cycle×number of tasks
output
ZW User's Manual