Known Problems

Software 1.6.0_C1 has following known problems:
Mammo test image has to be copied to D: drive before it is available for printing
on DT2 Mammo blue base films. Details see chapter 3.2, Tools and Auxiliary
means.
In very rare cases it is possible, that a false error message "MOD08 - please
replace module" is shown (MOD08 is the Devil CPU) after start-up of the printer.
A reset of the printer solves the problem.
The operator browser menu does not allow to delete images or set into
emergency status. It is possible however to delete print jobs in the operator menu
of the local keypad (refer to Drystar Axys user manual).
For TH profile calibration on a mammo printer a TH profile calibration film is
printed with too high density (approx. 2.0 O.D. instead of 1.2 O.D.). Entering
these higher values in the TH profile calibration browser screen is not possible,
as they exceed the upper limit. Using a Drystar DT2B or DT2C film allows to
perform the TH profile calibration properly.