AGFA DRYSTAR AXYS Service Manual page 270

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
2.5
Compliance Film Calibration as part of Afssaps / EUREF Calibration
Procedure
WHEN TO BE
The compliance film calibration may be required as last step of the Afssaps / EUREF*
PERFORMED
calibration procedure for a DRYSTAR AXYS used for printing mammo images.
The Afssaps / EUREF compliance calibration procedure has to be performed in
following cases:
*Afssaps
EUREF
1) The compliance film calibration is only
PREREQUISITE
available for DRYSTAR AXYS
software ≥ 1.6.2_C1.
2) A "TG18-PQC modified 4" image has
to be available on the modality.
The Test Pattern can be downloaded
from http://www.euref.org/.
Make film calibration this accurate, that it passes the acceptance criteria as defined
PURPOSE
by Afssaps / EUREF:
The default film calibration procedure is not necessarily accurate enough.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Before the Afssaps / EUREF mammo acceptance test.
Each 6 months, if the printer is under control of Afssaps / EUREF, and it did not
pass the Afssaps / EUREF calibration check (see section 2.5.6).
=
Agence française de sécurité sanitaire des produits de santé
=
European Reference Organisation for Quality Assured Breast Screening and
Diagnostic Services
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Figure 10: TG18-PQC modified 4
Agfa Company Confidential
Chapter 3.6 / 18