AGFA DRYSTAR AXYS Service Manual page 207

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC308002 : devilmodule.keypad.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC308003 : devilmodule.th.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
Cure3
Part
Reason4
Cure4
Part
IMPORTANT:
This fault may have following additional reason and cure:
Reason: Ambient temperature too low when switching on the printer.
Cure: Ensure that ambient temperature is within specifications (+15 ° C to +30 ° C)
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E308002
SeeManual
Connector P8 not connected
Connect P8
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
Defect
E308003
SeeManual
Connector P43 not connected
Connect P43
Connector P1 on TH not connected
Connect P1 on TH
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
TH defect
Replace
MOD05
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 41 of 72