DD+DIS033.07E
HC314021 : gearboxmodule.m8.shortcircuit
Type
Context
Action
Reason
Cure
Part
MODULE : suspensiondrivemodule
Elabel : No
HC315000 : suspensiondrivemodule.move
Type
Context
Action
Reason
Cure
HC315001 : suspensiondrivemodule.m3.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E314021
SeeManual
Roller drive module defect
Replace
MOD14
Accidental
None
Suspension move (10 or 14 inch) failed accidently
Self resolving accident
Defect
E315001
SeeManual
Motor M3 not connected
Connect motor M3
Suspension drive module defect
Replace
MOD15