AGFA DRYSTAR AXYS Service Manual page 272

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
Overview of the Afssaps / EUREF calibration procedure:
Drystar AXYS Mammo calibration procedure for Afssaps/EUREF
Check calibration of densitometer. Details see section 2.5.1.
Check configured default densitometer. Details see section 2.5.2.
YES
Reset CDM, TH, CFC calibrations. Details see section 2.5.3
Perform following calibrations. Details see section 2.5.4:
1. Film calibration for mammo trays (2 x)
2. CDM calibration
3. Film calibration for mammo trays (1 x)
4. Print Head calibration (1x)
5. Film calibration for mammo trays (2 x)
Perform density reproduction check (DRC). Details see section 2.5.5.
Print TG18 QPC pattern. Measure 18 density values.
Enter values in 'EUREF&AFSSAPS GSDF check.xls'. Details see 2.5.6.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Densitometer
calibration ok?
YES
Used densitometer
= default densitometer
in the printer?
YES
Printer installation date
= today?
No
DRC ok?
YES
Deviations < 10%?
YES
Finished
CFC_calibration_axys.cdr
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Adjustments and Calibrations
Perform densitometer
No
calibration acc. to densitometer
manufacturer instructions.
No
Adjust densitometer in 'Key-operator menu -
2 Change settings – 5 Image Quality
- 1 General settings'
No
No
Perform CFC calibration.
Details see section 2.5.7.
Repeat up to 3 times
if required.
Figure 11
Repair and Service
Chapter 3.6 / 20
Agfa Company Confidential