DD+DIS033.07E
MODULE : thmodule
Elabel : Yes
HC305000 : thmodule.high
Type
Context
Action
Reason
Cure
Part
HC305001 : thmodule.thpos.fail
Type
Context
Action
Reason
Cure
HC305002 : thmodule.thermalheadheater
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Reason3
Cure3
Reason4
Cure4
Reason5
Cure5
Reason6
Cure6
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E305000
SeeManual
Coolingfan defect
Replace
MOD16
Accidental
None
Internal jam
Resolve jam
Failure
W305002
SeeManual
Connector P3 on PSU not connected
Connect P3 on PSU
Connectors P2 and P3 on OSB not connected
Connect connectors P2 and P3
Connectors P6 on OSB not connected
Connect connector P6
Connector XS6 on heater not connected
Connect connector XS6
Sensor S12 defect
Replace MOD04
PSU defect
Replace