DD+DIS033.07E
HC308007 : devilmodule.p40.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC308008 : devilmodule.p1.connected
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC308009 : devilmodule.elabel.present
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E308007
SeeManual
Connector P40 not connected
Connect P40
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
Defect
E308008
SeeManual
Connector P1 not connected on MOD08 baseboard (Devil or Devius)
Connect P1 on MOD08 baseboard (Devil or Devius)
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
Failure
W308009
None
MOD08 baseboard (Devil or Devius) Elabel not connected
Connect MOD08 baseboard (Devil or Devius) elabel
MOD08 baseboard (Devil or Devius) Elabel defect
Replace
MOD08
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 43 of 72