DD+DIS033.07E
Cure3
Reason4
Cure4
Reason5
Cure5
Part
Reason6
Cure6
Part
Reason7
Cure7
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Connect connector P1 on TH
Connector P43 on MOD08 baseboard (Devil or Devius) not connected
Connect connector P43 on MOD08 baseboard (Devil or Devius)
PSU defect
Replace
MOD09
OSB defect
Replace
OSB
TH defect
Replace
MOD05