DD+DIS033.07E
(1)
Disconnect the printer from the
network.
(2)
Give the IP address of the
service PC the same IP address
like the printer*
(3)
Start ADVP
(4)
Select 'Basic Grayscale' in the
'Start up configuration' screen.
(5)
Select 'Application - Print
Emulator - Start'
(6)
In the following window fill in
- Called (Printer) AE Title (this is
the AE title of the printer),
- 'Called (Host) AE Title'
- ADVP Listen port (usually 104:
this is the port the modality
sends DICOM messages).
(7)
Select okay. Now ADVP is
prepared for recording the
communication between
modality and the printer.
NOTE:
*In case ADVP runs on the printer, just select 'Stop Controller' to stop the application
software (and so switch off the DICOM SCP functionality of the printer). When
communication recording is finished, close ADVP and select 'Start Controller'.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
Figure 52: ADVP as Print Emulator
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Tools and Auxiliary Means
Agfa Company Confidential
Chapter 3.2 / 51