2.5.6

Performing the Afssaps / EUREF Calibration Check

(1) Leave the service command session.
(2) Connect the printer network cable to the network outlet at the wall.
(3) Consult the "connectivity release document" for the modality connection to
DRYSTAR AXYS Mammo for the proper modality configuration to print the "TG18-
PQC modified 4" test image.
(4) Print the "TG18-PQC modified 4" test image from the modality.
(5) Measure the 18 density values and enter them in the Afssaps / EUREF
Calibration Check Excel sheet. See Figure 17.
EUREF/AFSSAPS GSDF check for printer
Laser:
Measured luminance values
testpattern
TG18-LN12-# p-value
1
0
2
240
3
480
4
720
5
960
6
1200
7
1440
8
1680
9
1920
10
2160
11
2400
12
2640
13
2880
14
3120
15
3360
16
3600
17
3840
18
4080
Result contrast response printer
1,000
0,100
0
(6) Check column "Deviation compared to DICOM (%)": All values have to be < 10%.
- If all values are < 10%, the procedure is finished.
- If one (or more) value(s) are ≥ 10 %, proceed with next section.
(7) Repeat this step for the other film tray, if the printer has two different DT2 Mammo
film formats loaded (e.g. 8x10" DT2 Mammo and 10x12" DT2 Mammo).
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
deviation compared
to DICOM (%)
Step 2: Check, that deviation is < 10% for all values. If 1 (or more) value > = 10% go to S
Luminance
JND
deviation compared
L
j(L)
dL/L
to DICOM (%)
1,798105
98,486
3,831783
142,848
0,722
9,0
7,638348
193,970
0,664
4,0
13,07981
241,359
0,525
3,5
21,51446
291,079
0,488
1,0
34,27055
342,479
0,457
3,9
51,35702
390,683
0,399
2,6
75,48349
439,328
0,380
1,8
111,1691
490,605
0,382
3,1
163,9521
544,108
0,384
7,2
231,176
592,847
0,340
2,3
333,7055
646,119
0,363
6,5
470,959
697,023
0,341
1,8
649,724
745,196
0,319
3,5
917,3471
797,368
0,342
4,3
1265,911
846,462
0,319
1,8
1747,063
895,837
0,319
1,3
2467,38
948,924
0,342
6,0
left monitor
DICOM 3.14
+/- 10%
1000
2000
3000
4000
5000
p-values
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Adjustments and Calibrations
DICOM 3.14 standard
testpattern
Luminance
JND
dL/L
TG18-LN12 p-value
L
j(L)
dL/L
+10%
1
0
1,80
98,486
2
240
4,17
148,511 0,794
0,874
3
480
8,07
198,537 0,638
0,702
4
720
14,11
248,563 0,544
0,599
5
960
23,09
298,589 0,483
0,531
6
1200
36,13
348,614 0,440
0,484
7
1440
54,76
398,640 0,410
0,451
8
1680
81,09
448,666 0,388
0,426
9
1920
117,99
498,692 0,371
0,408
10
2160
169,45
548,718 0,358
0,394
11
2400
240,90
598,743 0,348
0,383
12
2640
339,85
648,769 0,341
0,375
13
2880
476,63
698,795 0,335
0,369
14
3120
665,49
748,821 0,331
0,364
15
3360
926,13
798,846 0,328
0,360
16
3600
1285,79
848,872 0,325
0,358
17
3840
1782,24
898,898 0,324
0,356
18
4080
2467,38
948,924
0,322
0,355
lineair relation between p-value and JN
a: 0,208441
b: 98,4856
Figure 17
Type 5367/100
Repair and Service
Step 1: Enter the measured values he
dL/L
-10%
Step # on TG18 image Measured OD LuminanceStep # in CFC calibration screen on printer
1
3,7
1,7981049
0,715
2
3,15
3,8317831
0,575
3
2,78
7,6383476
0,490
4
2,52
13,079807
0,434
5
2,29
21,514455
0,396
6
2,08
34,270551
0,369
7
1,9
51,357016
0,349
8
1,73
75,483485
0,334
9
1,56
111,16915
0,322
10
1,39
163,95211
0,313
11
1,24
231,17597
0,307
12
1,08
333,70551
0,302
13
0,93
470,95902
0,298
14
0,79
649,72404
0,295
15
0,64
917,34706
0,293
16
0,5
1265,9111
0,291
17
0,36
1747,0633
0,290
18
0,21
2467,38
Step 3: Enter these values in CFC* screen on print
Only required if deviation >= 10% (see step 2)
* CFC = Compliance Film Calibration
Chapter 3.6 / 28