AGFA DRYSTAR AXYS Service Manual page 401

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS032.07E
Select densitometer:
Square
Measured density Target density Density diff. Check
100%
90%
80%
70%
60%
50%
40%
30%
20%
10%
0%
When entering the measured density values the decimal point appears in a different spot, e.g. measured value = 0.23, value to enter = 23.0
Density
Lower limit
0
30
45
60
75
90
105
120
135
150
165
180
195
210
225
240
255
270
285
300
315
NOTE:
The measured density values have to be entered without decimal point
(e.g. measured value = .23, value to enter = 23)
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
09-2007
X-rite 341
4
20
20
0
46
48,5
-2,5
74
77
-3
102
105,5
-3,5
130
134
-4
158
162,5
-4,5
186
191
-5
215
219,5
-4,5
244
248
-4
270
276,5
-6,5
300
305
-5
Upper limit
0
0
-3
3
-4
4
-5
5
-6
6
-7
7
-8
8
-9
9
-9
9
-9
9
-9
9
-10
10
-10
10
-10
10
-10
10
-10
10
-11
11
-11
11
-11
11
-11
11
-12
12
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Dmax = 305
0
Density Reproduction =
0
0
0
0
0
No MDM calibration needed for DT2B film
0
0
0
0
0
Density difference SMPTE image DT2B film DSAXYS
15
10
X-rite 310
X-rite 331
5
X-rite 341
X-rite 301
Mac Beth TR924
0
Gretag D200-2
-5
-10
-15
0
50
100
150
Target Density
Figure 20
Type 5367
Maintenance
OK
Lower limit
Upper limit
Density difference
200
250
300
350
Chapter 9 / 23
Agfa Company Confidential