DD+DIS033.07E
HC303021 : upperpickupmodule.m7.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC303022 : upperpickupmodule.m7.shortcircuit
Type
Context
Action
Reason
Cure
Part
HC303023 : upperpickupmodule.s14
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Failure
W303021
SeeManual
Upper pickup motor not connected
reconnect upper pickup motor
Upper pickup module defect
Replace
MOD03
Defect
E303022
SeeManual
Upper pickup module defect
Replace
MOD03
Failure
W303023
None
Upper pickup home sensor S14 malfunctioning
check upper pickup home sensor S14
Upper pickup module defect
Replace
MOD03
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 17 of 72