DD+DIS033.07E
HC309004 : powersupplymodule.powermotor.current
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC309005 : powersupplymodule.temperature.warning
Type
Context
Action
Reason
Cure
HC309020 : powersupplymodule.powerth
Type
Context
Action
Reason
Cure
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E309004
SeeManual
Actuators power cables short circuited
Check actuators power cables
PSU module defect
Replace
MOD09
Failure
W309005
None
PSU temperature high
Cool down the PSU
Defect
09
Replace
PSU module defect
Replace
MOD09