AGFA DRYSTAR AXYS Service Manual page 204

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC306027 : outputmodule.cdm.s10
Type
Context
Action
Reason
Cure
Part
HC306028 : outputmodule.cdm.working
Type
Context
Action
Reason
Cure
Part
MODULE : drumdrivemodule
Elabel : No
HC307001 : drumdrivemodule.m1.connection
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
IMPORTANT:
This fault may have following additional reason and cure:
Reason: Ambient temperature too low when switching on the printer.
Cure: Ensure that ambient temperature is within specifications (+15 ° C to +30 ° C)
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
06
Replace
Output module defect
Replace
MOD06
Defect
06
Replace
Output module defect
Replace
MOD06
Defect
E307001
SeeManual
M1 not connected
Connect M1
Drum drive module defect
Replace
MOD07
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 38 of 72