Software Patch 1.6.1_C1

Type of new Software*
Release Date
Printer Serial Numbers
Availability
Prerequisites
*NOTE:
A software upgrade is a functional extension. An update (software patch) is an
internal change designed to correct bugs, fix defects, or otherwise minor
improvements to performance without changing the software's basic design,
structure or functionality.
On MedNet, GSO Library always the latest available software release can be
downloaded.
2.1

Solved Software Bugs

With software patch 1.6.1_C1 following software bugs are solved:
Sporadic input film jams for films from the upper tray, mainly 14" films:
Symptom 1
Error message at the local display:
Error message in the infocounter:
The force for driving the film towards the thermal head is sporadically too low.
Cause
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
03-2011
Software Releases and Patches
Software Update
03/2008
SN  1188
MedNet, GSO Library
None
Film jam! Open upper input tray and
clear film path.