AGFA DRYSTAR AXYS Service Manual page 135

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
3.5.3.1
Modules Pages
Main Purpose
Sub menus
Comments:
1.
By pressing activate/deactivate the motor can be started
2.
To check a sensor, activate the sensor manually and press "REFRESH". Now the
changed state of the sensor has to be displayed, if the sensor is still okay
3.
It is possible to view the E-label content of the corresponding module.
Refer to Figure 23.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
10-2007
Figure 21: Service Engineer - Modules Pages
Allows checking of individual motors and light sensors of
the printer.
Note: This function is only for special cases. Normally the
device detects each defective motor and sensor and
issues a corresponding error code.
By clicking on the assembly, the corresponding assembly
page is displayed. See Figure 22
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Tools and Auxiliary Means
Agfa Company Confidential
Chapter 3.2 / 29