DD+DIS033.07E
HC309001 : powersupplymodule.elabel.present
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC309002 : powersupplymodule.elabel.config
Type
Context
Action
Reason
Cure
Part
HC309003 : powersupplymodule.powerth.current
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
Reason3
Cure3
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Failure
W309001
None
PSU Elabel not connected
Connect PSU elabel
PSU module Elabel defect
Replace
MOD09
Failure
W309002
None
PSU module Elabel defect
Replace
MOD09
Defect
E309003
SeeManual
TH power cables short circuited
Check TH power cables
TH module defect
Replace
MOD05
PSU module defect
Replace
MOD09