Electrical Check According To National Regulations - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS032.07E
(1)
Insert the power cord at the printer
side; leave it disconnected at the
wall side.
(2)
Put one probe at the ground
connector of the mains connector.
(3)
Put the other probe at a few metallic
parts, the customer could touch.
The electrical resistance in this case
has to be ≤ 0.3 Ohm.
Refer to Figure 15.
IMPORTANT:
If the electrical resistance of one measurement exceeds the limits (0.2 Ohm without
mains cable; 0.3 Ohm with mains cable) check the printer for loose grounding
connections!
3.9.2

Electrical Check according to National Regulations.

(1)
NOTE:
The electrical checks that have to be performed in certain intervals (e.g. each 12
months) or after certain service actions (repair and/or maintenance) depend on
National Regulations.
Example for Germany: BGV A3 and DIN VDE 0751 or subsequent regulations.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 0
09-2007
Perform the electrical check according to National Regulations.
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367
Maintenance
536631JN.cdr
Figure 15
Chapter 9 / 18
Agfa Company Confidential