DD+DIS033.07E
5.4.2

Image Adjustments as described in the Connectivity Release Document

(1) If a Connectivity Release Document for the Printer - Modality Combination exists,
perform the image adjustments as described there.
(2) When finished, print test images from the modality.
(3) Check the image quality together with the customer.
(4) In case the image quality is not satisfying continue as described in the flowchart
(see Figure 15
NOTE:
Connectivity Release Documents describe all parameters and required adjustments to
guarantee:
Communication between modality and the printer
Image adjustment at the modality and / or the printer.
MEDNET
Connectivity Release Documents are available in MEDNET, HealthCare Library -
Path
General Info - Connectivity & Application.
5.4.3

Image Adjustments at the Modality

The Image adjustments at the modality are split up in contrast (see 5.4.3.1
sharpness adjustments (see 5.4.3.2
5.4.3.1
Contrast Adjustment
(1) Select the required taste LUT (Look Up Table) at the modality. If possible use the
"Kanamori Taste LUT".
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Controls, Connections and Setup Procedure
).
X
NOTE:
The Look-Up-Table is responsible for the contrast of the image (in
combination with Dmin and Dmax) and may be named at the modality as
follows:
Taste LUT
Perception LUT
Correction LUT
Presentation LUT.