AGFA DRYSTAR AXYS Service Manual page 211

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC308013 : devilmodule.power.n12v
Type
Context
Action
Reason1
Cure1
Part
Reason2
Cure2
Part
HC308014 : devilmodule.power.24v
Type
Context
Action
Reason1
Cure1
Part
Reason2
Cure2
Part
HC308015 : devilmodule.power.actuator
Type
Context
Action
Reason1
Cure1
Part
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E308013
SeeManual
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
PSU module defect
Replace
MOD09
Defect
E308014
SeeManual
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
PSU module defect
Replace
MOD09
Defect
E308015
SeeManual
MOD08 baseboard (Devil or Devius) module defect
Replace
MOD08
PSU module defect
Replace
MOD09
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 45 of 72