AGFA DRYSTAR AXYS Service Manual page 177

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC302007 : lowertraymodule.s2
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC302016 : lower.filmcal.required
Type
Context
Action
Reason
Cure
HC302018 : lowerTag-unsupported
Type
Context
Action
Reason
Cure
HC302023 : backside.jam.lower
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Failure
W302007
None
S2 not connected
Connect S2
Lower tray module defect
Replace
MOD02
Failure
W302016
None
Lower tray needs calibration
Perform lower tray calibration
Accidental
None
Wrong version in lower tray
Replace filmpack
Accidental
None
backside jam while printing from lower tray
Insert filmpack correctly