AGFA DRYSTAR AXYS Service Manual page 286

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
medicalimage test image.
Perform Resistor Check.
* Service Engineer Tools - Enter Service Command Session -
3 Print Testimage – 1 Test image - STI2.ZIP
** browser service menu- 7 Test & Diagnose –
6 Diagnose mechanics – 5 Print Head Unit – 3 Resistor Check
*** Service Engineer Tools - Enter Service Command session
select '6. Calibration - 3. Print Head Calibration
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
TH cleaning
procedure
*
Print STI2 and
No
Vertical stripes
visible ?
Yes or in doubt
Clean TH with
Isopropyl alcohol
("wet" cleaning).
*
Print STI2 test image.
Additional
vertical stripes
visible ?
No
No
Previous Stripes
still visible ?
Yes
**
Resistor Check okay ?
Yes
Perform Dry TH
Cleaning Procedure.
IMPORTANT:
Observe TH cleaning rules
described next page.
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Adjustments and Calibrations
Okay. No
replacement
required
Refer to figure next page.
Yes
Okay. No
replacement
required.
No
Replace TH.
536736mm.cdr
Figure 20
Type 5367/100
Repair and Service
Dry TH cleaning
procedure
Dismount TH.
Clean TH with
Isopropyl alcohol
("wet" cleaning).
Use ink eraser
for cleaning.
Clean TH with
Isopropyl alcohol
("wet" cleaning).
Mount TH.
Clean TH with
Isopropyl alcohol
("wet" cleaning).
*
Print STI2 test image.
New vertical
stripes visible ?
No
Vertical
No
stripes visible ?
Yes
***
Perform TH
profile calibration.
Stripes visible in
No
medical_image test
image ?
Yes
Replace TH.
Chapter 3.6 / 34
Agfa Company Confidential
Yes
Okay. No
replacement
required.
Okay. No
replacement
required.