Resetting The Compliance Film Calibration, Internal Densitometer Calibration And Thermal Print Head Calibration; Performing Internal Densitometer Calibration, Film Calibrations And Thermal Print Head Calibration - AGFA DRYSTAR AXYS Service Manual

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
2.5.3
Resetting the Compliance Film Calibration, Internal Densitometer Calibration and
Thermal Print Head Calibration
NOTE:
Resetting the Compliance Film Calibration, Internal Densitometer Calibration and
Thermal Print Head Calibration is required to be sure, that possibly wrong calibrations
from previous service interventions are deleted.
(1) In the service command session select "1 Show settings – 2 Input Tray".
(2) Check, which tray is using "DT2M" film material.
(3) In the service command session select "6 Calibration – 6 Compliance film
calibration – 1 Reset compliance film calibration for upper tray or 2 Reset
compliance film calibration for lower tray". (depends on which film tray is using
DT2M film material).
If both trays contain DT2M film material, select both trays, one after another.
(4) Leave the "service command session"
(5) In the service tools page select "Reset calibration data"
(6) Select "Default CDM" and "Default Profile". Note, that "Default Profile" button is not
available, if no "Print Head Profile" has been performed in the field yet.
(7) Leave the Reset calibration data" screen.
NOTE:
After this action the printer will show warnings
HC302016 Lower tray needs film calibration - or
HC301016 Upper tray needs film calibration - and
HC306014 Printer needs CDM calibration
This problem will be solved in the next section.
2.5.4
Performing Internal Densitometer Calibration, Film Calibrations and Thermal
Print Head Calibration
NOTE:
The Internal Densitometer Calibration, Film Calibration and Thermal Print Head
Calibration is described in this chapter (3.6 Adjustments and Calibrations) in detail.
Here only a short description is given. For details refer to ...
2.2 CDM (Internal Densitometer) Calibration
2.1 Film Calibration
2.3 Print Head (TH) Calibration
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Chapter 3.6 / 22
Agfa Company Confidential

Advertisement

Table of Contents
loading

Table of Contents