AGFA DRYSTAR AXYS Service Manual page 174

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
HC301007 : uppertraymodule.s1
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
HC301016 : upper.filmcal.required
Type
Context
Action
Reason
Cure
HC301018 : upperTag-unsupported
Type
Context
Action
Reason
Cure
HC301019 : tag.history.insecure
Type
Context
Action
Reason
Cure
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Failure
W301007
None
S1 not connected
Connect S1
Upper tray module defect
Replace
MOD01
Failure
W301016
None
Upper tray needs calibration
Perform upper tray calibration
Accidental
None
Wrong version in upper tray
Replace filmpack
Defect
E301019
SeeManual
Tag history is corrupt
make Emergency sw install without preserving config
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Troubleshooting - Error Codes
Agfa Company Confidential
Chapter 3.3 / 8 of 72

Advertisement

Table of Contents
loading

Table of Contents