DD+DIS033.07E
Accessible via ...
Accessible via menu:
Purpose:
Default value:
Example browser
screen:
NOTE:
• The Called AE title is defined during the execution of the printer installation wizard.
It can also be adapted in the "DICOM Print SCP Setup" screen.
• The number of logical SCP printers is limited. As each configured SCP server
consumes resources of the printer, use as little SCP servers as possible. Rule of
thumb: Define max. 4 logical printers.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Controls, Connections and Setup Procedure
Browser only.
Browser pages - Service Engineer Tools - DICOM profile
editor - Setup DICOM print SCP Servers
Define more than one logical printer. This may be required in
case a modality sends images from different applications to
the printer. See also Info next page.
Only one logical printer.
Internal name: DEFAULT_SCP
Called AE title: DSAXYS
HOME HELP
DICOM Print SCP Setup
Setup one or more Dicom SCP servers on this device:
One 'Dicom Print SCP server' is capable of handling multiple Dicom
print associations
Dicom Print SCP
Internal Name:
DEFAULT_SCP
Called AE title:
DSAXYS
Listen at Port:
104
Max. allowed associations:
Secure channel (SSL):
Comments:
OK
Create new Dicom Print SCP
DOCUMENT CONTROL NOTE: