DD+DIS033.07E
MODULE : lowertraymodule
Elabel : No
HC302000 : lowertraymodule.connected
Type
Context
Action
Reason
Cure1
HC302001 : lowertray.overruled
Type
Context
Action
Reason
Cure
HC302002 : lowertray.disabled
Type
Context
Action
Reason
Cure
HC302003 : lowertraymodule.tagreader
Type
Context
Action
Reason1
Cure1
Reason2
Cure2
Part
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2011
Defect
E302000
SeeManual
XS9 not connected to lower tray module
Connect XS9 to lower tray module
Failure
W302001
None
Inputtray overruled (dmax 2.0)
Replace filmpack
Failure
W302002
None
Inputtray disabled
Replace filmpack
Failure
W302003
None
Tag reader not connected
Connect tag reader
Tag reader defect
Replace
Tag reader