AGFA DRYSTAR AXYS Service Manual page 282

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
(13) Leave the "service command session".
(14) Repeat step "2.5.6 Performing the Afssaps / EUREF Calibration Check.
IMPORTANT:
If after three repetitions one or more "Deviation compared to DICOM (%)" values are
still ≥ 10%, start the complete procedure from scratch. Refer to section 2.5.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
If this time all "Deviation compared to DICOM (%)" values are < 10%, the
"Afssaps / EUREF calibration procedure" is finished.
If one (or more) "Deviation compared to DICOM (%)" value(s) are ≥ 10%,
repeat steps 2.5.6 and 2.5.7 up to three times. See IMPORTANT note
below.
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Type 5367/100
Repair and Service
Adjustments and Calibrations
Agfa Company Confidential
Chapter 3.6 / 30