AGFA DRYSTAR AXYS Service Manual page 258

Type 5367/100
Hide thumbs Also See for DRYSTAR AXYS:
Table of Contents

Advertisement

DD+DIS033.07E
via local display
Measure left
white border
(1)
On the local keypad select key-operator menu '3 Print Image - 1 Testimage –
Verification:
upper or lower tray - SMPTE_xxx.tif': An SMPTE test image will be printed.
(2)
Measure the left and right blank frame in print direction: It must be
Result:
The horizontal film alignment procedure is performed.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
06-2008
Left white border
0.0 mm
Measure right
white border
0.0 mm
o ≥ 1,5 mm for DT2B / DT2C Films
o ≥ 1 mm for DT 2 Mammo Films.
DOCUMENT CONTROL NOTE:
DRYSTAR AXYS
Adjustments and Calibrations
Measure the right clear border width:
Measure the left clear border width:
Value:
mm
Value:
mm
Service - Calibration
Left white border
Right white border
Accept changes
Figure 1
Type 5367/100
Repair and Service
Right white border
via Browser User Interface:
Welcome to this
DryStar AXYS
Horizontal alignment
0.0
(mm)
0.0
(mm)
Discard changes
Chapter 3.6 / 6
Agfa Company Confidential
536736ac.cdr