2.5.6

Performing the Afssaps / EUREF Calibration Check

(1) Leave the service command session.
(2) Connect the printer network cable to the network outlet at the wall.
(3) Consult the "connectivity release document" for the modality connection to
DRYSTAR 5302 for the proper modality configuration to print the "TG18-PQC
modified 4" test image.
(4) Print the "TG18-PQC modified 4" test image from the modality.
(5) Measure the 18 density values and enter them in the Afssaps / EUREF
Calibration Check Excel sheet. See Figure 17.
EUREF/AFSSAPS GSDF check for Genrad printers
Laser:
Measured luminance
testpattern
TG18-LN12-# p-value
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
Result contrast response printer
(6) Check column "Deviation compared to DICOM (%)": All values have to be < 10%:
- If all values are < 10%, the procedure is finished.
- If one (or more) value(s) are ≥ 10 %, proceed with next section.
(7) Repeat this step for the other film tray.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
05-2009
deviation compared
to DICOM (%)
Step 3: Check, that deviation is < 10% for all values. If 1 (or more) value > = 10% go to Step 4
values
Luminance
JND
deviation compared
L
j(L)
dL/L
to DICOM (%)
0
11,98582
233,211
240
17,90569
272,089
0,396
2,3
480
25,06399
307,434
0,333
6,5
720
34,43093
343,017
0,315
5,6
960
47,83903
382,004
0,326
3,3
1200
67,27169
424,518
0,338
12,0
1440
90,89841
463,665
0,299
3,0
1680
121,6709
502,866
0,290
3,0
1920
157,2109
538,238
0,255
6,8
2160
199,6444
571,900
0,238
11,1
2400
260
609,777
0,263
0,0
2640
339,5642
648,671
0,265
2,7
2880
454,5699
691,754
0,290
13,5
3120
582,7169
728,841
0,247
2,2
3360
731,0079
762,970
0,226
9,9
3600
938,8381
800,887
0,249
0,1
3840
1206,575
839,125
0,250
0,9
4080
1551,488
877,616
0,250
1,5
1,000
left monitor
DICOM 3.14
+/- 10%
0,100
0
1000
2000
3000
4000
5000
p-values
DOCUMENT CONTROL NOTE:
DRYSTAR 5302
Adjustments and Calibrations
Step 2: Adjust Illumination and refected light used for printing the ima
"Illumination cd/m
2500
"reflected (ambient) light"
10
DICOM 3.14 standard
testpattern
Luminance
JND
TG18-LN12 p-value
L
j(L)
dL/L
1
0
233,211
11,99
2
240
17,74
271,117 0,387
3
480
25,43
309,023 0,356
4
720
35,61
346,929 0,333
5
960
48,95
384,835 0,316
6
1200
66,33
422,741 0,301
7
1440
88,83
460,648 0,290
8
1680
117,87
498,554 0,281
9
1920
155,23
536,460 0,274
10
2160
203,16
574,366 0,268
11
2400
264,57
612,272 0,263
12
2640
343,13
650,179 0,259
13
2880
443,52
688,085 0,255
14
3120
571,73
725,991 0,253
15
3360
735,41
763,897 0,250
16
3600
944,32
801,803 0,249
17
3840
1210,97
839,709 0,247
18
4080
877,616
0,247
1551,49
lineair relation between p-value and JN
a: 0,157942
b: 233,2106
Figure 17
Type 5366/100
Repair and Service
Step 1: Enter the measured values here
e
2"
dL/L
dL/L
+15%
-15%
Step # on TG18 image Measured OD LuminanceStep # in CFC calibration screen on printer
1
3,1
11,98582
0,445
0,329
2
2,5
17,905694
0,410
0,303
3
2,22
25,06399
0,383
0,283
4
2,01
34,430931
0,363
0,268
5
1,82
47,839031
0,347
0,256
6
1,64
67,271691
0,334
0,247
7
1,49
90,898414
0,323
0,239
8
1,35
121,6709
0,315
0,233
9
1,23
157,21091
0,308
0,227
10
1,12
199,64439
0,302
0,223
11
1
260
0,297
0,220
12
0,88
339,56418
0,294
0,217
13
0,75
454,56985
0,290
0,215
14
0,64
582,71691
0,288
0,213
15
0,54
731,00788
0,286
0,211
16
0,43
938,83807
0,285
0,210
17
0,32
1206,5752
0,284
0,210
18
0,21
1551,4875
Step 4: Enter these values in CFC* screen on print
Only required if deviation >= 10% (see step 2)
* CFC = Compliance Film Calibration
Chapter 3.6 / 27
Agfa Company Confidential
1 7
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
0

Advertisement