Increasing- Load Test (Phase 2)

NOTE
Product Reference Guide
Figure 5-3. European Shift Test (Metric)
After completing the Shift Test, you must complete the Increasing Load
Test using 10.00, 12.50 and 15 kilograms of weight.
The upper limit of the scale is configurable according to POS interface type and
may not necessarily be set at 15 kilograms, which is the standard setting. For this
test, continue to place weights in 2.50 kilgram increments only up to the upper
weight limit set for your scale.
1. Place 10.00 kilograms in the center of the weighing surface and
check that the display reads between 9.990 kg and 10.010 kg.
2. Place an additional 2.50 kilograms in the center of the weighing sur-
face and check that the display reads between 12.490 kg and 12.510
kg.
3. Place an additional 2.50 kilograms in the center of the weighing sur-
face and check that the display reads between 14.990 kg and 15.010
kg.
4. Remove the weights and verify that the display reads 0.000 kg.
5. You have completed phase two of the increasing load test.
Calibration Verification (Kilograms)
A
B
CENTER
C
D
5-13