Overview Of Diagnostic Information - Endress+Hauser Proline Prosonic Flow I 400 Operating Instructions Manual

Ultrasonic time-of-flight flowmeter modbus rs485
Hide thumbs Also See for Proline Prosonic Flow I 400:
Table of Contents

Advertisement

Proline Prosonic Flow I 400 Modbus RS485
Endress+Hauser
12.8

Overview of diagnostic information

The amount of diagnostic information and the number of measured variables affected
increase if the measuring device has one or more application packages.
In the case of some items of diagnostic information, the diagnostic behavior can be
changed. Adapting the diagnostic information →  134
Diagnostic
Short text
number
Diagnostic of sensor
019
Device initialization
active
082
Data storage
inconsistent
083
Memory content
inconsistent
104
Sensor signal path 1 to n 1. Check process conditions
105
Downstream transducer
path 1 to n defective
106
Upstream transducer
path 1 defective
160
Signal path switched off
Diagnostic of electronic
201
Electronics faulty
242
Firmware incompatible
252
Module incompatible
262
Module connection
interrupted
270
Main electronics
defective
271
Main electronics faulty
272
Main electronics faulty
273
Main electronics
defective
275
I/O module defective
276
I/O module faulty
Remedy instructions
Device initialization in progress, please
wait
Check module connections
1. Restart device
2. Restore S-DAT data
3. Replace S-DAT
2. Clean or replace transducers
3. Replace sensor electronic module
(ISEM)
1. Check connection to the
downstream transducer
2. Replace downstream transducer
1. Check connection to the upstream
transducer
2. Replace upstream transducer
Contact service
1. Restart device
2. Replace electronics
1. Check firmware version
2. Flash or replace electronic module
1. Check electronic modules
2. Check if correct modules are
available (e.g. NEx, Ex)
3. Replace electronic modules
1. Check or replace connection cable
between sensor electronic module
(ISEM) and main electronics
2. Check or replace ISEM or main
electronics
1. Restart device
2. Replace main electronic module
1. Restart device
2. Replace main electronic module
Restart device
1. Pay attention to display emergency
operation
2. Replace main electronics
Change I/O module
1. Restart device
2. Change I/O module
Diagnosis and troubleshooting
Status
Diagnostic
signal
behavior
[from the
[from the
factory]
factory]
S
Warning
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
M
Warning
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
F
Alarm
1)
1)
135