Replace The Fuses - Leica THERMOBRITE ELITE User Manual

Table of Contents

Advertisement

7.6.3

Replace the Fuses

Turn the main power switch located in the back panel to the Off (0) position. Unplug the ThermoBrite Elite
Unit from the wall outlet and disconnect it from the Netbook before replacing the fuses.
For continued protection against risk of fire and hazard, replace only with the same type and rating fuses.
1. The instrument fuses are located on the rear panel, between the power cable receptacle and the power switch.
2. Remove the power cable to access the fuse holder.
3. Using a small flat screwdriver on one side tab, pry the fuse holder out.
ThermoBrite Elite User Manual
3855-7500-501 B02
Figure 7-13: Power Switch—Fuse Box
Figure 7-14: Power Switch—Access Fuse Holder
7. Maintenance
97

Advertisement

Table of Contents
loading