Replacing The Fuses - Leica CM1510S Instruction Manual

Cryostat
Table of Contents

Advertisement

8.
Cleaning and maintenance
8.7

Replacing the fuses

F4
F5
F6
F7
F8
Fig. 31
1
2
Fuse
Protects:
F4
Elreha controller
F5
Defrost valves 1 and 2
F6
Pipe defrost heater
F7
Heating of condensate
water collection tray
F8
Sliding window heater
46
Before replacing the fuses, turn off the instrument mains switch
and pull the mains plug!
Only use replacement fuses of the same specification as the
original ones. For the required values, please refer to chapter 3.2
'Technical Data'!
Fuses other than the ones specified can cause severe damage to
the instrument.
Replacing the fuses
On the back of the instrument there is a fuse board with 5 fuses:
• Insert a screw driver into the slot in the fuse cap; lightly press the
fuse inwards, then rotate 1/4 turn to the left and let go. The fuse will
automatically be released and can then be easily removed.
• Remove defective fuse (1) from cap (2) and insert a replacement fuse
of identical specification.
• Put the fuse cap back in, lighty press inwards with the screw driver
and fasten rotating 1/4 turn to the right.
Type
T 0.5 A
T 1.0 A
T 2.0 A
T 4.0 A
T 2.0 A
Instruction manual V 1.4 – 04/2010