Replacing Main Fuses - Leica HistoCore SPECTRA CV Instructions For Use Manual

Coverslipper
Table of Contents

Advertisement

8
Malfunctions and Troubleshooting
8.4

Replacing main fuses

Warning
The instrument should be shut down using the
the case of a instrument failure. The main fuses can then be inspected.
A suitable flat-head screwdriver must be used to prevent damage to the fuse holders.
Beware of defective fuses! Potential risk of injury due to broken glass! Wear suitable protective
clothing (safety goggles, cut-resistant gloves
1. To do so, open the hood and unscrew the two fuse holders
cover with a flat-head screwdriver
2. Check the fuse being used for damage.
3. Remove the defective fuse from the holder and replace it with a new fuse.
4. Installation occurs in the reverse sequence of operations.
152
Main Switch
(→ P. 18 – 3.1 Standard delivery – packing
1
Fig. 126
and disconnected from the power supply in
(→ Fig. 126-1)
on the top of the right
Version 2.1, Revision K
list)).