6.

Cleaning and Maintenance

6.1

Cleaning the instrument

• Before cleaning the instrument, turn off using the power switch
from the power supply.
• The instrument has a continuous front panel and no lateral cooling slots, so the surfaces
are smooth and easy to clean.
• All surface materials are resistant to common laboratory detergents.
The instrument must be switched off and have cooled down for cleaning.
Do not use organic solvents (e.g. alcohol or xylene)!
6.2

Replacing fuses

Switch the unit off and unplug it. Use
only the supplied replacement fuses.
Both fuses must have the same rat-
ing (check the imprint!). Fuses must
always be switched out in pairs.
• Insert a small screwdriver into the cutout
(Fig.
8, 11) and carefully pry out the fuse
insert (12).
• Remove the fuse holder and replace the
defective fuses with two new ones (13).
• Then reinsert the fuse holder and press it
into place.
14
(Fig.
6, 3) and disconnect
11
13
Instructions for Use V 3.4 RevI – 09/2018
12
Fig. 8