7. Maintenance
6. Insert a new pump tube into the inlet and exit plugs. Push the tube all the way to avoid any possible leak.
7. Guide the tubing into the peristaltic pump, above the rollers. Make sure the tube rests in the V notch on both sides
of the pump. See image below.
8. Lower the Peristaltic Pump Cover to close the pump. This lowers the top part of the pump, securing the tube
above the rollers.
9. Discard the old tube according to applicable regulations.
10. Repeat the procedure for the second peristaltic pump.