6

Cleaning and maintenance

6.
Cleaning and maintenance
6.1

Cleaning the instrument

• Before cleaning the instrument, turn it off using the main switch and disconnect it from the power
supply.
• Empty the water tray.
• The instrument must have cooled down for cleaning.
• Use wet, lint-free tissue paper to clean the instrument.
• Cleaning products suitable for the removal of paraffin can be used to clean the instrument.
• Do not use organic solvents (e.g. alcohol or xylene).
6.2

Replacing fuses

Warning
Replacing the fuses without switching off the instrument and unplugging the power plug.
Electrical shock which causes injury of persons.
• Turn the instrument off with the power switch and disconnect the power plug from the mains before
replacing the fuses. Empty the water tray.
Caution
Using the wrong fuses which are not with the same specification defined in the section of Technical
data in the Instructions for Use.
Instrument malfunction.
• Only use fuses of the same specification defined in the section of Technical data in the Instructions for
Use.
Use only the supplied replacement fuses. Both fuses must have the same rating (check the imprint).
Replace the fuses in pairs.
Fig. 18
1. Use a straight screwdriver
direction.
2. Pull out the fuse holder
3. Replace the defective fuses with two new ones.
4. Insert the fuse holders and use the straight screwdriver to rotate them in place.
24
2
(→ Fig. 18-1)
to rotate the fuse holder
(→ Fig. 18-3)
carefully.
1
(→ Fig. 18-2)
to 10 o'clock
Version 1.0, Revision A
3